top of page
Date | Name | Affiliation | Other Observers | Instrument | Hours Observed | Hours Lost | Reason | Observing Program |
|---|---|---|---|---|---|---|---|---|
10/02/2021 | Martha Leake | Valdosta State University | Alicia Heger | ARC camera | 0 | 12 | Student projects and asteroid light curves. I was unable to connect to KPNO. Gary said that he could get it and that the problem may be local. Using RADMIN, I was not able to connect with any of the three KPNO computers, although I could connect to CTIO and ORM. I will be checking with our IT folks soon. Skies were partly cloudy. Xianming was able to use the night. | |
10/01/2021 | XIANMING HAN | Butler Univ | Josh Wanninger | CCD | 10 | 0 | eclipsing binary and asteroid.
ACE helped fixing the camera (which was reported not functioning in the previous day). | |
09/30/2021 | XIANMING HAN | Butler University | Josh Wanninger | CCD | 0 | 10 | Lost first half of the night due to clouds. Lost 2nd half due to camera issues:
The CCD camera on SARA KP is not functioning. It does take images. But all images look like bias. I tried different exposure times, different filters, checked X-stage guider positions and made sure the spectrograph is not checked, turning on and off the dome red lights, telescope mirror cover open and close. But none of these seem to make any difference.
Have emailed ACE. And tried to call ACE but did not succeed yet. | |
09/29/2021 | William Keel | University of Alabama | 2 cats | ARC CCD | 0 | 12 | Dead quasar survey, wiped out by clouds thicker and lasting longer than in the forecast. | |
09/27/2021 | Keefe Kamp | Florida Tech | Norton, Gokuldass, Tobolski, Baranowski | CCD and Spec | 0 | 10 | Be Star and TESS follow-up observations planned. Right at twilight threatening clouds came in and did not look like they were going anywhere. | |
09/24/2021 | Kyle Pellegrin | Clemson University | Samalka Anandagoda, Dr. Hartmann | ARC CCD | 0 | 5 | Couldn't observe due to clouds and chance of rain during the night. | |
09/23/2021 | Kyle Pellegrin | Clemson University | Samalka Anandagoda, Dr. Dieter Hartmann | ARC CCD | 4 | 0 | Observed starburst galaxies. | |
09/21/2021 | XIANMING HAN | Butler Univ | Sera Shaughnessey | CCD | 5.5 | 4.5 | CCD camera froze. Tried to access Switch 1 to recycle power to ACE control Cabinet. But switch 1 is not accessible (only switch 2 is accessible). Will email ACE about this problem. After restarting ACE software, a camera does begin to take images, but this camera appears to be a different camera (same stars do not show up). So please take a look. | |
09/20/2021 | Gary Henson | ETSU | none | Spectrograph | 9.5 | 0 | Spectroscopy of Massive Stars, Spectrograph appears to be working nominally after summer shutdown.
FLI ccd cooled to -37C running TEC at about 97% power. With external air temp ~17C, focus settings were Fiber plate (-515), Spectrum (-3650), Telescope (151000).
Found track rates of 15.065, +0.011 to be reasonably good before autoguiding, but autoguiding was excellent and system stable although I observed and tracked a single target all night. | |
09/18/2021 | Keefe Kamp | FIT | Caballero, Gokuldass | CCD, Spec | 0 | 12 | Planed Be photometric and Spectroscopic observations however with a 50% chance of rain tonight. decided not to open. | |
09/16/2021 | Tomomi Otani | ERAU | Brian Herbster | CCD | 0 | 10 | Since dome control UPS appears to have failed, Gary suggested us not to open the dome if precipitation threat within 24 hours. Precipitation threat tonight and tomorrow, so we could not open the dome. | |
09/16/2021 | Todd Vaccaro | Ball State University | None | ARC CCD | 8.5 | 1 | Eclipsing binary stars. Imaging sequence froze with ccd computer stuck at 100% exposure and tele computer unable to take exposure; after finding the instructions for resetting things on the camera computer I lost about half an hour. Another half hour was lost at the end of the night as tracking failed at about HA 3:30 and I missed the last half of eclipse egress. |